Home | About Fylde | Applications | Products | Data Sheets | Contact Us | Search

Micro Analog 2 - FE-MM40

FE MM40 Menu ItemFE MM16 Menu ItemFEMM8 menu itemFE M4 menu itemFEMA32 menu itemModules menu itemConnectors menu itemUSB

The FE-MM40 is a DC powered 40 channel combined signal conditioning and data acquisition system. Like all chassis styles in the Micro Analog 2 range, it accepts a range of dual channel signal conditioning modules which together cover almost every transducer type. There are slots for up to 20 dual channel modules. An additional slot is reserved for the optional FE-366-AZ auto zero module which controls the automatic offset removal for FE-366-TA bridge transducer amplifiers. Inputs are via the 7 pin rear panel connectors which can accept a variety of connector adaptors. A 50 way connector allows a 40 way BNC adaptor to be connected (FE-MAC-40C). This allows single ended +/-10V signals to be input without using signal conditioning, or allows the output of the signal conditioning modules to be routed to other measuring equipment.

This system is used with the USB interface by inserting the FE-357-USBM interface module.

The front panel drops down to allow any of the Micro Analog 2 modules to be inserted into any position. Configuration and operation is simple, and the rear panel connector may be connected to a 40 way BNC adaptor or connected using an adaptor to any data acquisition system.

FEMM40 Front ViewFE-MM40 system

20 input connectors (one for each pair of channels) are shown in the rear panel illustration.

FEMM40 Rear ViewFE-MM40 system rear panel

The front panel drops down to reveal the signal conditioning modules.

FE-MM40 system with panel down

Applications

Load Cells, Pressure Transducers, Strain Gauges, DCLVDTs, Potentiometers, all forms of bridge type transducers and DC voltage signals, Accelerometers, Hydrophones, ICP® transducers and AC voltage signals, Thermocouples, Frequency Pickups. Mix and Match cards as required.

System Specification

Maximum 40 channel analogue instrumentation system in 2U enclosure. System can be supplied with rack mounting ears or front panel handles.

Up to 20 dual channel cards in any combination plus optional Auto-Zero card.

10 to 36 V DC supply (Max current 2 A).

Inputs 20 x 7 pin Tuchel. Output 50 way SCSI1 syle connector. Adaptors available for both input and output connectors.

Dimensions 100(H) x 440(W) x 375(D)

Environment 0°C to 40°C, 0 to 80 % R.H. non condensing.

 